6-6
BP2 Setup Factory Defaults
BP2 SETUP ADULT PED NEO DESCRIPTION
ALARMS OFF OFF OFF Enables BP2 alarm
ALARM LEVEL 2 2 2 Sets alarm priority level (1, 2, or 3)
LATCH ALARMS OFF OFF OFF Latches the BP2 alarm until reset
SYS HI LIMIT 50 40 20 Systolic high alarm limit setting
SYS LO LIMIT 20 5 5 Systolic low alarm limit setting
DIA HI LIMIT 25 15 10 Diastolic high alarm limit setting
DIA LO LIMIT 5 00 00 Diastolic low alarm limit setting
MN HI LIMIT 30 25 15 Mean high alarm limit setting
MN LO LIMIT 10 5 5 Mean low alarm limit setting
PRAM AVAIL ON ON ON BP2 enable or disable
PRAM ON / OFF OFF OFF OFF BP2 standby
RCD ON ALARM WF WF WF Record when BP2 alarm enabled
SIZE 30 30 30 BP2 waveform size
SCALE OFF OFF OFF BP2 scale enable
DISPD VALUES S/D S/D MEAN BP2 numeric display format
SITE OFF OFF OFF BP2 catheter site
PRAM COLOR YELLW YELLW YELLW Sets on-screen color of parameter