1-10
1. External DC Input (12 - 28V) Connector
Allows an external DC power source to be connected for extended use during transport where AC
power is not available. Batteries are also charged with external DC power.
2. Fuse Holder and Fuse Replacement
The fuse holder contains the fuses for the monitor. Open the door of this holder to replace fuses.
It is recommended that qualified technical service personnel replace fuses when necessary.
WARNING: For protection against fire, replace the fuses only with
those of the same type and rating.
The Prism SE requires two identical slow blow T2.5A / 250V fuses for operation. Ensure that the
AC power cord has been disconnected before replacing fuses. Carefully open the door of the fuse
holder with a short 1/8" flat screwdriver. Replace the blown or defective fuse with one of the
same type and rating. Noting orientation, gently slide the fuse holder back until it locks snugly
into its original place. After the fuse replacement, connect the AC power cord to the Prism SE.
The unit is now ready for operation.
3. AC Power Cord Input Connector and Optional Power Cord Mounting Bracket
Allows connection of the AC power cord to the monitor. In addition, the AC power cord must be
plugged into an AC source before the monitor can operate on AC power and before the batteries
can be charged.
For permanent connection, the MDE power cord bracket should be installed. See the following
figure for instructions on installing the power cord mounting bracket.