Escort Prism 20401 Dokumentacja Strona 52

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 51
3-11
T1 WAIT. DO NOT place probe into position until message reads T1 READY and an
audible tone is heard.
3. When ready, place probe in position. ORAL and RECTAL methods require caregiver to
hold probe in position.
4. An audible tone for NORM mode signals completion of the measurement. A message on
screen displays T1 FINAL. The MONIT mode is continuous monitoring of temperature so
no audible tone will occur to signal a final temperature. The message on screen notes
T1 MONITOR.
5. Probe covers are discarded after use by “ejecting” probe cover off probe via blue/red ejects
button.
6. Replacement of probe into probe well is required to “reset” device.
WARNING: Incorrect insertion of rectal probe can cause bowel
perforation. Read all instructions thoroughly before proceeding with
probe placement.
NOTE: Regarding the use of SureTemp:
Do not initiate a temperature while unit is displaying T1 INIT on
screen (when monitor is first activated)
Correct positioning of probe tip is critical with oral method
NORM mode allows predictive temps of 4 seconds for oral site and
15 seconds for rectal site
Axillary site should be assessed in MONITOR mode
Probe should be reseated in probe well between temperatures to
allow recalibration of parameter
If ambient temperature of monitor is above 90
o
F, or if the body
temperature is below 94
o
F, then NORMAL mode cannot be used
and the monitor will automatically switch to MONITOR mode
Noninvasive Blood Pressure Measurements (NIBP)
Prism and Prism SE noninvasive blood pressure (NIBP) monitoring can be started at any time by
pressing the NIBP START/STOP key on the monitor’s front panel, or is user-configurable to take
measurements at preset intervals. NIBP measures oscillometric waveforms as the air pressure in a
standard cuff gradually reduces.
Additionally, initial and subsequent cuff inflations are adaptable for different patient types and
blood pressure ranges. Depending on which patient mode is selected (ADULT, PED, or NEO),
initial inflations can be setup anywhere from 60 to 220 mmHg, and subsequent readings will
automatically inflate to only 25 mmHg above the previous systolic reading to increase
measurement speed and patient comfort.
Przeglądanie stron 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag