There is arterial occlusion proximal to sensor
Patient is in cardiac arrest or is in shock
SPO2 Sensor Accessories—Masimo
Name/Type Model Number Application
LNOP
- ADT LNOP
- ADT
Adult, digit
LNOP
- PDT LNOP
- PDT
Adult/Pediatric
LNOP
- NEO LNOP
- NEO
Neonatal < 10 kg
LNOP
- NEO PT LNOP
- NEO PT
Neonatal < 1 kg
LNOP
- DC I LNOP
- DC I
Adult/Pediatric
PC04 Patient Cable: 4 Ft. PC04 Adult/Pediatric/Neonatal
PC08 Patient Cable: 8 Ft. PC08 Adult/Pediatric/Neonatal
PC12 Patient Cable: 12 Ft. PC12 Adult/Pediatric/Neonatal
WARNING: Incorrect sensor application or use may cause tissue
damage or improper operation of the monitor. Follow directions for
use provided with sensor.
NOTE: Regarding use of LNOP
sensors:
Before use, carefully read LNOP
sensor directions for use.
Use only Masimo oximetry sensors for SPO2 measurements. Other
oxygen transducers (sensors) may cause improper performance.
Do not use damaged LNOP
sensors or patient cables. Do not use
an LNOP
sensor with exposed optical components. Do not
immerse sensor or patient cable in water, solvents, or cleaning
solutions (sensors, cable, and connectors are not waterproof). Do
not sterilize by irradiation, steam, or ethylene oxide. See cleaning
instructions in Directions for Use for reusable Masimo LNOP
sensors and patient cables.
SPO2 Alarm Delay
The SPO2 alarm incorporates a delay to minimize nuisance alarms related to brief erroneous
SPO2 measurements that can be caused by patient artifact or external noise. The factory default
time for this delay is 5 seconds and, if desired, a 10-second delay is also available by accessing
the ALM DELAY softkey under the SPO2 setup menu. The monitor will generate an alarm if the
averaged SPO2 numeric value exceeds the alarm limit for more than 5 or 10 seconds, depending
on the delay selected.
SPO2 Averaging Time
SPO2 measurements are taken every 2 seconds and are averaged over a period of time (the
designated averaging MODE) to display the most accurate oxygenation percentage and pulse
rate. The MODE key in the SPO2 setup page lets you select a mode of measurement averaging to
best fit your clinical needs.