3-23
UNIT
%
mmHg
O2
ON
OFF
N2O
ON
OFF
ETCO2 allows numeric information and waveform to be displayed. ETCO2 waveform can be
placed in a trace slot. Refer to configuration instructions.
CAUTION: Do not over tighten FilterLine tubing in connector.
CAUTION: The Microstream CO2 FilterLine tubing is for use with
approved airway adapters only.
CAUTION: CO2 airway adapter and FilterLine are non-sterile and
intended for single patient use only.
CAUTION: Do not re-circulate exhaust gas back into closed loop
breathing system.
WARNING: Do not attempt to clear or “blow out” the FilterLine.
Replace FilterLine when occluded.
To measure a patient’s ETCO2 (intubated):
1. Attach airway adapter proximally to patient between endotracheal, nasotracheal or tracheal
tube and ventilator circuit, T-tube setup or bag/valve setup. CO2 airway adapter is non-sterile
and intended for single patient use only.
2. Connect appropriate end of ETCO2 FilterLine tubing to Prism SE’s connector panel. CO2
airway adapter and FilterLine are non-sterile and intended for single patient use only.
3. Observe WARM UP message on screen (WARM UP appears 5 seconds after a NO
SENSOR message). ETCO2 monitoring will commence approximately 10 seconds following
setup and connection.
To measure a patient’s ETCO2 (nonintubated):