6-1
Chapter 6 – Factory Defaults
When the Prism & Prism SE monitor is powered on for the first time, it will be in the Neonatal
mode, and all parameters will be set to their factory defaults.
Each parameter has defaults for adult, pediatric, and neonatal modes (ADULT, PED, NEO). All
parameter and system defaults can be user-configured.
It is important to recognize that changes made to these functions determine the status of those
functions upon every subsequent power-up. As a result, default settings should not be changed
unless the setting will be desired on a regular basis.
It is recommended that you have service personnel perform any factory default/deep
configuration functions.
System Setup Factory Defaults
SYSTEM SETUP ADULT PED NEO DESCRIPTION
FACTORY DEFAULTS NO NO NO To restore all settings to factory defaults
PWR DN MEMORY 15MIN 15MIN 15MIN Selected time for saving VSC settings
ALRM VOLUME 7 7 7 Alarm tone volume setting (1-8)
ALM VOL MIN 1 1 1
Sets minimum allowable alarm tone
volume setting (1-8)
TONE VOLUME 4 4 4 Key tone setting
TRACE 2 AUTO AUTO AUTO Waveform or feature for second trace
TRACE 3 AUTO AUTO AUTO Waveform or feature for third trace
TRACE 4 AUTO AUTO AUTO Waveform or feature for fourth trace
HR SOURCE AUTO AUTO AUTO Heart rate source, ECG, SPO2, BP’s
ALM SUSPND 180 90 60 Alarm suspend time
PRAM LABELS BLANK BLANK BLANK To blank parameters when off
CLOCK DISPLY OFF OFF OFF Time displayed on screen
CABLE / TELM CABLE CABLE CABLE ECG mode - telemetry or cable
REMOTE RECD OFF OFF OFF Telemetry recording
ALARM FLASH ON ON ON Flashing alarm LED
LANGUAGE ENGL ENGL ENGL
Sets language (English, French,
German, Spanish, Swedish, or Italian)
LINE FREQ 60HZ 60HZ 60HZ 50Hz or 60Hz input line frequency
PATIENT TYPE ALL ALL ALL ADULT, PED, NEO, or ALL
ALARM CNTRL ON ON ON
When default is set to ON, allows
enabled alarms to be turned OFF