9-14
2. Press the CIS softkey associated with the ENTER VITAL SIGNS function. The
following CIS menu will be displayed with the corresponding CIS softkeys shown
below. NOTE: the following values depict the default values.
RESP: 12
TEMP: 98.6 (37.0)
NIBP: 120 / 80 (95)
SPO2: 95
PR: 95
3. Enter manual vital sign data using softkeys shown below.
NEXT
ADJ
ADJ
F4 F5
NEXT ↓→ —Moves cursor to the next parameter.
ADJ —Increases value.
ADJ —Decreases value.
QUIT—Exits the manual entry mode without saving values.
STORE—Stores all valid values into the VSC table.
NOTE: To facilitate the entry of manual vital signs, the first time
the ADJ key is pressed, the last saved parameter value will appear.
If no prior value has been stored, the parameter default value will
be displayed. Upon selection of the NEW PATIENT function, all
parameter values will return to their default values.
4. Press the STORE softkey to save manually entered values in the patient record.
CIS Module Storage Capacity
The CIS Communication Module has the ability to store unsent patient vital signs information
collected for a given patient. This feature is useful in the event of a network failure or when there
is a need to store vital signs data for several patients prior to the ability to confirm the patients
with the hospital’s ADT system. The vital sign data is stored directly in the CIS Module which
contains its own backup battery such that all data is retained even when the Patient Monitor is
turned off. A total of 2500 entries for up to 10 patients can be stored in the CIS module. Please
see section “Special Circumstances”, for details on how to send vital sign data stored in the CIS
module to the Clinical Information System.
Deleting Vital Signs Entries
The CIS Communication Module has the ability to allow the deletion of vital signs entries from
the VSC table and CIS. When an entry is deleted from the VSC table, that entry will no longer
appear in the patient record at the Clinical Information System.
To delete an entry from the VSC table:
Przeglądanie stron 131