Przeglądanie stron 133
9-16
NOTE: If the network connection is lost per the above situation, the
CIS Status Message at the top of the monitor will no longer simply
display the confirmed patient name. Instead, the patient name will
flash and alternate with the message “NO CIS FOUND”. This will
not only alert you to the fact that your network connection is lost,
but will also indicate the patient the CIS Module is saving vital sign
data for in the absence of the network connection.
Situation 2:
Inability to Complete the Patient Confirmation Process Prior to Measuring Vital Signs for a
New Patient.
CIS STATUS MESSAGE:
“SETUP PATIENT” and Temp Patient Name (Alternating and Flashing)
Under certain circumstances, it is possible that the patient confirmation process may not be
completed prior to your need to monitor a different patient with that same monitor. One example
of these potential circumstances could include the hospital network being down during the entire
monitoring period for the prior patient – which of course precludes the ability to confirm the
patient. Another scenario that could prohibit the completion of the confirmation process prior to
monitoring a subsequent patient is if the patient has not been entered into the hospital’s ADT
system.
As indicated previously, vital sign data can only be sent to the CIS after the patient has been
confirmed via the monitor. In the situations described above where this confirmation cannot be
completed, each CIS Module is capable of storing a total of 2500 vital sign entries for up to (10)
different patients. This saved vital sign data can be sent to the CIS once the confirmation process
can be completed for any patient whose data is stored in the CIS Module. The key to successfully
storing vital sign data in the CIS Module for multiple patients is assigning a patient name for each
patient – even though circumstances preclude the operator from confirming the patient with the
ADT system. The CIS Module permits the operator to enter the patient’s name for this purpose
(referred to as a “temporary” name in the Monitor’s CIS menu) as described in detail below.
NOTE: When the user has entered a “temporary” name, the CIS
Status Message at the top of the monitor will flash this temporary
name and alternate with the message “Setup Patient”. This should
alert you to the fact that the patient is NOT confirmed while also
indicating the patient name the CIS Module is saving vital sign data
for, prior to the completion of the confirmation process.
Przeglądanie stron 133