Sample Waveform Packet
Packet ID
2 bytes
Waveform
Number
1 byte
Length
1 byte
Data
Check
Sum
1 byte
1b 57 00 13 83 83 81 80 7d 98 b3 cb b3 96 7a 7e 80 81 83 83 83 83 83 7e
Table 3: Parameter Indices
Parameter # Waveform # Parameter Name
0 0 ECG
1 1 Respiration
2 BP1 Systolic
3 2 BP1 Diastolic
4 BP1 Mean
5 BP2 Systolic
6 5 BP2 Diastolic
7 BP2 Mean
8 NIBP Systolic
9 Not Available NIBP Diastolic
10 NIBP Mean
11 11 SaO
2
12 Not Available Temperature
13 Not Available CO - Escort II packets only
14 Not Available CO
2
- Escort II packets only
15 15 ECG Vector 2 (No Parameter – Waveform Only)
NOTE: Temperature is transmitted as an integer value with a range
between 200 and 500 (Base 10). Each integer represents one-tenth
degree Celsius (0.1° C). This translates to an actual temperature
range transmitted between 20 and 50 degrees Celsius
(20-50° C).
NOTE: There is no parameter block for ECG Vector 2 although it is
listed as “Parameter and Waveform #15”. This is simply an
additional waveform associated with ECG Parameter and Waveform
#0.
Table 4: Parameter Status
Hex
Value
ASCII
Value
Parameter Status
4Eh ‘N’ Parameter Connected, Alarms ON
46h ‘F’ Parameter Connected, Alarms OFF
41h ‘A’ Parameter Connected, in Alarm
44h ‘D’ Parameter Disconnected
56h ‘V’ Parameter not Available
5Ah ‘Z’ Parameter not Zeroed